CASE REPORT
Year : 2016  |  Volume : 43  |  Issue : 1  |  Page : 30-33

Button battery ingestion: A therapeutic dilemma and clinical issues in management


1 Department of Radiodiagnosis, Adesh Hospital, Kurushetra, Haryana, India
2 Department of Anaesthesiology and Intensive Care, Adesh Hospital, Kurushetra, Haryana, India
3 Department of Anaesthesiology and Intensive Care, Mullana Medical College, Solan, Himachal Pradesh, India
4 Department of Medical Health Services, Patiala, Punjab, India

Date of Web Publication2-Feb-2016

Correspondence Address:
Sukhminder Jit Singh Bajwa
House No. 27-A, Ratan Nagar, Tripuri, Patiala - 147 001, Punjab
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-5009.175454

Rights and Permissions
  Abstract 

Button batteries, also known as disc batteries, are tiny, round batteries commonly present in numerous household electronic gadgets such as watches, calculators, cameras, hearing aids, penlights, remote control devices, and certain toys. Because of their easy accessibility to children, they pose a serious menace of accidental ingestion or inhalation. Most of the ingestions are benign. However, large-sized button batteries when ingested can become impacted in the esophagus or elsewhere in the gastrointestinal tract with life-threatening repercussions. We report a 4-year-old female child brought to our setup by her parents with an alleged history of ingestion of a button battery of the watch approximately about 2 h back. The child, apart from being anxious, was asymptomatic. An initial radiograph of the chest and abdomen was taken, which showed the battery at the distal end of the stomach. Considering the small size of the battery cell and after seeking the gastroenterologist's consultation, it was decided that the child would be managed conservatively with oral antacids and serial abdomen skiagrams were obtained. The button battery was eventually detected in the stools after 2 days.

Keywords: Button battery, impaction, ingestion


How to cite this article:
Bhangu JS, Bajwa SS, Anand S, Kalsi K. Button battery ingestion: A therapeutic dilemma and clinical issues in management. J Sci Soc 2016;43:30-3

How to cite this URL:
Bhangu JS, Bajwa SS, Anand S, Kalsi K. Button battery ingestion: A therapeutic dilemma and clinical issues in management. J Sci Soc [serial online] 2016 [cited 2022 Aug 17];43:30-3. Available from: https://www.jscisociety.com/text.asp?2016/43/1/30/175454


  Introduction Top


Button batteries are ubiquitous, now being increasingly found in household electronic gadgets such as watches, calculators, hearing aids, penlights, remote control devices, and certain toys. Their smooth and shiny appearance coupled with their easy accessibility has inadvertently led to their frequent ingestion or inhalation by children. [1] A majority of the battery cell ingestions are benign, passing spontaneously through the gastrointestinal tract. [2] However, they can get lodged in the esophagus or elsewhere in the gastrointestinal tract where they can be fatal and life-threatening. [3],[4] These button cells, apart from being ingested, can be inhaled by placing them in the nose or they can be placed in the ear as well. [1] In fact, button battery ingestion accounts for 2% of all the foreign body ingestions. [5],[6] Ingestion of large-sized button batteries (≥20 mm) in children younger than 4 years is associated with increased morbidity and mortality [7] While a small-sized button battery passes uneventfully through the gut within 2-6 days, sometimes taking up to 2-4 weeks, [8] large-sized button cells can get impacted in the gut, the most common site being the esophagus. [9] An impacted button cell can ulcerate, perforate, lead to fistula formation, or even death. [10]


  Case report Top


A 4-year-old female child was brought to our setup with an alleged history of ingestion of button cell battery of a watch 6.8 × 1.1 mm in diameter (as mentioned in the other battery cell of the same watch) about 2 h back. On examination the child was conscious, cooperative, oriented but anxious. Her weight was 15 kg. Her heart rate was 110 beats per minute, blood pressure was 90/60 mmHg, and saturation by pulse oximeter was 98%. Her abdomen was soft, nondistended, and nontender. Bowel sounds were normal. Radiographs of the chest and abdomen were taken, which showed the button cell at the distal end of the stomach [Figure 1]. Gastroenterologist consultation was sought and considering the small size of the cell, its presence in the distal stomach, and asymptomatic presentation, the child was kept on conservative treatment. The child was admitted and kept on normal diet. Syrup ranitidine 2.5 mL (35 mg) once daily was started. Also, serial skiagrams of the abdomen were taken [Figure 2] and [Figure 3]. The stool was also examined for the cell. The child remained asymptomatic with stable vitals and the button cell was eventually detected in the stool after 2 days.
Figure 1: Skiagram of day 1 showing button battery in the distal part of the stomach

Click here to view
Figure 2: Day 2 image showing movement of button battery to the large gut

Click here to view
Figure 3: Abdominal x-ray of day 3 showing no sign or image of button battery

Click here to view



  Discussion Top


Button battery ingestion poses a serious menace to children. The first reported case of button cell ingestion was in 1977 involving a child who had swallowed a camera battery, which had become impacted in the proximal esophagus. [11] Now because of its easy accessibility, there has been an alarming surge in the accidental ingestion, especially in the very young.

Button cell batteries are mainly of four types: Mercury, silver, lithium, and a strong hydroxide of sodium or potassium. [12] Lithium cells are commonly present in watches. Lithium cells though have a higher potential of 3 V as compared to 1.5 V seen in other cell types but being thin, they can easily pass through the gut if swallowed. [2] The mechanism of button battery injury includes generation of an external electrolytic current that hydrolyzes tissue fluids and generates hydroxide at the anode, leakage of battery contents and their absorption, low voltage burns, and pressure necrosis of the adjacent tissue. [8],[13] Button cells can result in considerable morbidity and mortality once impacted in the esophagus or elsewhere in the gut within a few hours of its ingestion, thereby emphasizing the need of distinguishing it from other foreign body ingestions. Complications include esophageal burns and perforations, tracheoesophageal fistulas, mediastinitis, and aortic perforation. [4] Chest and abdominal skiagrams with appropriate exposure can prove to be very useful in these cases. Button cells give a characteristic double rim appearance in the anterior-posterior view and step-off appearance in the lateral view. [4]

In the present case, the child presented to us approximately 2 h after ingestion of the button cell of the watch. At the time of presentation, the child was anxious but asymptomatic with stable vitals. The child had swallowed a single battery cell of the watch, which had a dimension of 6.8 × 1.1 mm (as mentioned on the other battery cell of the same watch). Immediate skiagrams of the chest and abdomen were obtained, which showed the presence of the disc cell at the distal end of stomach. As the child was asymptomatic and the battery cell had already crossed the esophagus, it was decided that the child would be managed conservatively on a normal diet and syrup ranitidine was started. Though the child could be managed at home as well, since she belonged to a distant village it was decided that she would be admitted.

It has been found that virtually all the button cells that cross the lower esophageal sphincter will pass out without any hurdle, thereby negating the need for endoscopic retrieval. [2] Emesis should not be evoked at this time as there is a risk of the theoretical model of regurgitation and fatal airway obstruction. [2],[14] The administration of oral antacids has been found to be useful as they decrease corrosion of the cell. [15] Small cells as in the present case, usually negotiate the pylorus without any need for endoscopic retrieval or surgical intervention. However, serial abdominal radiographs are required to confirm the passage. Rarely can a cell get impacted in the Meckel's diverticulum. [16] Endoscopy or surgical removal has a role to play only if the cell gets impacted in the esophagus, stomach, or elsewhere in the gut mucosa for prolonged periods or the signs of peritonitis set in. [17] Ingestion of multiple battery cells with ≥20 mm dimension by a child of less than 4 years age are associated with adverse outcomes. [7] Symptoms of button cell impaction in the decreasing order of frequency include vomiting or hematemesis, pain in the abdomen, discoloration of stool, fever, loose stool, rashes, respiratory distress, and nonspecific symptoms such as anorexia, dysphagia, and retrosternal discomfort. [18] As none of the above symptoms were present in the present case and the cell had passed the esophagus, a decision for conservative management was taken. The child was closely monitored for the above mentioned symptoms; serial skiagrams were taken and inspection of the stool was done. The battery cell was detected in the stool on the second day.

Our case report highlights that in cases where the battery cell has been ingested and found to be present in the stomach with the patient being asymptomatic, an overzealous treatment is not required. Rather it is safe to consider a conservative approach in management. However, a check on the impending signs and symptoms should be kept and definitive intervention carried out accordingly.


  Conclusion Top


Button battery cell ingestions have shown an alarming surge in the recent years among children because of their easy accessibility. Prompt medical attention is required in such cases. Though button cells lodged in the esophagus require urgent endoscopic intervention, those present in the stomach at the time of presentation should be managed conservatively in the asymptomatic child coupled with serial follow-up abdominal skiagrams and stool examination. The button cells are usually detected in the stool within 2-6 days. [8] Definitive management in the form of endoscopic retrieval or surgical intervention may be required in case of development of impending signs and symptoms. However, prevention is still the need of the hour. Parents should be educated about the potential hazards of the battery cell exposure, especially ingestion among children and the need to keep these away from their reach. Also, manufacturers should ensure secure packaging of these cells.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Lin VY, Daniel SJ, Papsin BC. Button batteries in the ear, nose and upper aerodigestive tract. Int J Pediatr Otorhinolaryngol 2004;68:473-9.  Back to cited text no. 1
    
2.
Litovitz TL. Battery ingestions: Product accessibility and clinical course. Pediatrics 1985;75:469-76.  Back to cited text no. 2
    
3.
Hamilton JM, Schraff SA, Notrica DM. Severe injuries from coin cell battery ingestions: 2 case reports. J Pediatr Surg 2009;44:644-7.   Back to cited text no. 3
    
4.
Yardeni D, Yardeni H, Coran AG, Golladay ES. Severe esophageal damage due to button battery ingestion: Can it be prevented? Pediatr Surg Int 2004;20:496-501.   Back to cited text no. 4
    
5.
Litovitz T, Schmitz BF. Ingestion of cylindrical and button batteries: An analysis of 2382 cases. Pediatrics 1992;89:747-57.   Back to cited text no. 5
    
6.
Sheikh A. Button battery ingestions in children. Pediatr Emerg Care 1993;9:224-9.  Back to cited text no. 6
    
7.
Litovitz T, Whitaker N, Clark L, White NC, Marsolek M. Emerging battery-ingestion hazard: Clinical implications. Pediatrics 2010;125:1168-77.  Back to cited text no. 7
    
8.
Temple DM, McNeese MC. Hazards of battery ingestion. Pediatrics 1983;71:100-3.  Back to cited text no. 8
    
9.
Kimball SJ, Park AH, Rollins MD 2 nd , Grimmer JF, Muntz H. A review of esophageal disc battery ingestions and a protocol for management. Arch Otolaryngol Head Neck Surg 2010;136:866-71.  Back to cited text no. 9
    
10.
Chang YJ, Chao HC, Kong MS, Lai MW. Clinical analysis of disc battery ingestion in children. Chang Gung Med J 2004;27:673-7.  Back to cited text no. 10
    
11.
Blatnik DS, Toohill RJ, Lehman RH. Fatal complication from an alkaline battery foreign body in the esophagus. Ann Otol Rhinol Laryngol 1977;86:611-5.   Back to cited text no. 11
    
12.
Banerjee R, Rao GV, Sriram PV, Reddy KS, Nageshwar Reddy D. Button battery ingestion. Indian J Pediatr 2005;72:173-4.  Back to cited text no. 12
    
13.
Samad L, Ali M, Ramzi H. Button battery ingestion: Hazards of esophageal impaction. J Pediatr Surg 1999;34:1527-31.  Back to cited text no. 13
    
14.
Litovitz TL. Button battery ingestions. A review of 56 cases. JAMA 1983;249:2495-500.  Back to cited text no. 14
    
15.
Litovitz T, Butterfield AB, Holloway RR, Marion LI. Button battery ingestion: Assessment of therapeutic modalities and battery discharge state. J Pediatr 1984;105:868-73.  Back to cited text no. 15
    
16.
Wills GA, Ho WC. Perforation of Meckel′s diverticulum by an alkaline hearing aid battery. Can Med Assoc J 1982;126:497-8.  Back to cited text no. 16
    
17.
David TJ, Ferguson AP. Management of children who have swallowed button batteries. Arch Dis Child 1986;61:321-2.  Back to cited text no. 17
    
18.
Slamon NB, Hertzog JH, Penfil SH, Raphaely RC, Pizarro C, Derby CD. An unusual case of button battery induced traumatic tracheoesophageal fistula. Pediatr Emerg Care 2008;24:313-6.  Back to cited text no. 18
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]


This article has been cited by
1 Çocuklarda Yabanci Cisim Yutulmasinda Büyük Tehlike: Piller
Sebahat ALTUNDAG,Zehra ÇALISKAN
Saglik Bilimleri Dergisi. 2020;
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case report
Discussion
Conclusion
References
Article Figures

 Article Access Statistics
    Viewed6445    
    Printed117    
    Emailed0    
    PDF Downloaded321    
    Comments [Add]    
    Cited by others 1    

Recommend this journal